1MCP Process/ISMATEC SA/29.05.07/CB/GPEnglish FrançaisDeutschBetriebsanleitungOperating ManualMode d‘emploiPumpen-AntriebMikroprozessor-gesteuertSchut
10MCP Process/ISMATEC SA/29.05.07/CB/GP Start-up informationAfter switching on the power sup-ply switch, the following values are displayed:1 LED
11MCP Process/ISMATEC SA/29.05.07/CB/GPOperating panel1 Digital LED display 2 Reduce value * settings3 Increase value4 RUN/STOP – Starts and stop
12MCP Process/ISMATEC SA/29.05.07/CB/GP mmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVolumePROGRAMFlow rateTOTAL Anzeige der BetriebsartLEDs for operating
13MCP Process/ISMATEC SA/29.05.07/CB/GPbaMODEmmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVolumePROGRAMFlow rateTOTAL Control keysa RUN/STOP Starts and stops
14MCP Process/ISMATEC SA/29.05.07/CB/GP settingsokdefaultabcdesettingsGrundeinstellungenBasic settingsRéglage de base*Control keys for basic setti
15MCP Process/ISMATEC SA/29.05.07/CB/GPSelecting the basic settings1 Select the program (p. 20)2 Switch the pump off3 Keep the settings key pressed
16MCP Process/ISMATEC SA/29.05.07/CB/GP Program specific basic settingsk Individually adjustable for each of the 4 programs.CyclesNumber of dispens
17MCP Process/ISMATEC SA/29.05.07/CB/GPProgram specific basic settingsk Individually adjustable for each of the 4 programs.Foot switch– Toggle (default
18MCP Process/ISMATEC SA/29.05.07/CB/GP okokokokokokMenu Grundeinstellungen verlassenQuit the menue basic settingsQuitte le menu des réglages de b
19MCP Process/ISMATEC SA/29.05.07/CB/GPPeristaltik-PumpenkopfPeristaltic pump head Tête péristaltiqueKolben-Pumpenkopf Piston pump headTête pistoniØ i
2MCP Process/ISMATEC SA/29.05.07/CB/GP ContentsSafety precautions 4–5Warranty terms 6Product 7Rear panel 8Mains voltage 8Starting the pump 9
20MCP Process/ISMATEC SA/29.05.07/CB/GP Program selectionWhen switching the pump on, it always selects the previously used program.1 Change mode
21MCP Process/ISMATEC SA/29.05.07/CB/GPPumping by drive speed1 Change mode to PUMP rpm, 1.0 – 240.0 rpm, adjustable in steps of 0.1 rpm2 – Enter r
22MCP Process/ISMATEC SA/29.05.07/CB/GP Pumping by flow rate1 Change mode to PUMP Flow rate 2 Enter the required flow rate with 56 keys (is disp
23MCP Process/ISMATEC SA/29.05.07/CB/GPCalibrating the flow rate1 Change mode to PUMP Flow rate Enter the required flow rate with the 56 keys2 C
24MCP Process/ISMATEC SA/29.05.07/CB/GP Dispensing by time The dispensing time can be en-tered from 0.10 s to 999 h.1 Change mode to DISP Time2
25MCP Process/ISMATEC SA/29.05.07/CB/GPDispensing by volume1 Change mode to DISP Volume2 Use the 56 keys for entering the required dispensing volum
26MCP Process/ISMATEC SA/29.05.07/CB/GP Calibrating the volume1 Change MODE to DISP Volume k Use the 56keys for entering the required
27MCP Process/ISMATEC SA/29.05.07/CB/GPMODE1mmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVolumePROGRAMFlow rateTOTAL CALmmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVol
28MCP Process/ISMATEC SA/29.05.07/CB/GP mmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVolumePROGRAMFlow rateTOTAL mmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVolume
29MCP Process/ISMATEC SA/29.05.07/CB/GPIntermittent dispensing (by time)Intermittent dispensing by time with a pre-set pause time1 Go to mode PAUSE T
3MCP Process/ISMATEC SA/29.05.07/CB/GPContentsDispensing by volume within a pre-set time 28Intermittent dispensing by time 29 by volume 30No.
30MCP Process/ISMATEC SA/29.05.07/CB/GP mmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVolumePROGRAMFlow rateTOTAL mmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVolume
31MCP Process/ISMATEC SA/29.05.07/CB/GPmmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVolumePROGRAMFlow rateTOTAL mmrpmTimeTimeTUBE I.D. PUMPDISPPAUSEVolumePROGR
32MCP Process/ISMATEC SA/29.05.07/CB/GP Drip-free dispensingReverse roller steps (backsteps) allow drip-free dispensing (1 to 100 roller steps). W
33MCP Process/ISMATEC SA/29.05.07/CB/GPPumping against pressureThe MCP Process can be used for continuous duty at a differential pressure of max. 1.5
34MCP Process/ISMATEC SA/29.05.07/CB/GP Overcurrent protectorThe drive MCP Process features an overload protector. When an over-load condition occ
35MCP Process/ISMATEC SA/29.05.07/CB/GPHinweisWir verweisen auf unsere ausführli-che Schlauch-Dokumentation.Please notePlease refer to our detailed tu
36MCP Process/ISMATEC SA/29.05.07/CB/GP Analog interfacePin 1, GND (ground)Reference potential for all other inputsPin 2, remoteFor changing betwe
37MCP Process/ISMATEC SA/29.05.07/CB/GPAnalog interfacePin 9, speed OUTThe default setting is 0 – 10 VDC, proportionally to the motor speed 0 – 240.0
38MCP Process/ISMATEC SA/29.05.07/CB/GP Réglages du switch S1Commutateur coulissant S2Cet interrupteur influence le pin 9, speed OUTPosition A: 0
39MCP Process/ISMATEC SA/29.05.07/CB/GPSerial interfaceRS232 IN (female)A 9-pin D-socket is available on the rear panel of the pump.Setting the pump a
4MCP Process/ISMATEC SA/29.05.07/CB/GP Safety precautionsISMATEC® Pumps are designed for pumping applications in laborato-ries and industry. As su
40MCP Process/ISMATEC SA/29.05.07/CB/GP Serial interfaceRS232 OUT (male)Pin 2: RS232 RxIs used for connecting additional pumps together with pin
41MCP Process/ISMATEC SA/29.05.07/CB/GPSerielle Schnittstelle / Serial interface / Interface sérielle Befehl Funktion / Beschreibung Beispiel
42MCP Process/ISMATEC SA/29.05.07/CB/GP Serielle Schnittstelle / Serial interface / Interface sérielleBefehl Funktion / Beschreibung Beispiel A
43MCP Process/ISMATEC SA/29.05.07/CB/GPSerielle Schnittstelle / Serial interface / Interface sérielleBefehl Funktion / Beschreibung Beispiel Antwor
44MCP Process/ISMATEC SA/29.05.07/CB/GP Serielle Schnittstelle / Serial interface / Interface sérielleBefehl Funktion / Beschreibung Beispiel A
45MCP Process/ISMATEC SA/29.05.07/CB/GPSerielle Schnittstelle / Serial interface / Interface sérielleBefehl Funktion / Beschreibung Beispiel Antwor
46MCP Process/ISMATEC SA/29.05.07/CB/GP Serielle Schnittstelle / Serial interface / Interface sérielleBefehl Funktion / Beschreibung Beispiel A
47MCP Process/ISMATEC SA/29.05.07/CB/GPJede Pumpe braucht eine eigene Adresse. (siehe Seite 39).Each pump requiresan individual address. (see page 39)
48MCP Process/ISMATEC SA/29.05.07/CB/GP Logiciel de programmationLogiciel ProgEdit Créez votre programme d'application individuel. Chargez
49MCP Process/ISMATEC SA/29.05.07/CB/GPAccessoiresCarte d’interface Moxa C168incluant un câble multiple avec fiches 25 pôles (mâle) pour com-mander un
5MCP Process/ISMATEC SA/29.05.07/CB/GPSafety precautions Do not manipulate the pump head before the pump is swit-ched off and disconnected from the
50MCP Process/ISMATEC SA/29.05.07/CB/GP Cassettes de rechangeCassettes de rechange en POM-C CA Click'n'go cassette IS 3710CA levier de
51MCP Process/ISMATEC SA/29.05.07/CB/GP2-Stopper-Schläuche2-stop collared tubingTubes à 2 arrêtsAccessoiresMatériau POM-CCopolymère de polyoxyméthylèn
52MCP Process/ISMATEC SA/29.05.07/CB/GP MontageJedem Pumpenkopf liegt eine Mon-tageanleitung bei.Vor Inbetriebnahme den korrekten ID-Code eingeben
53MCP Process/ISMATEC SA/29.05.07/CB/GPedoCDI*.oNDI.rN-lletseB.oNredrOednammocedoNnetarßeilFsetarwolFtibéDnim )nim/lm( xamkcurDnoisserPerusserP)rab(sl
54MCP Process/ISMATEC SA/29.05.07/CB/GP epyT/pyT epyT/pyTepyT/pyTepyT/pyT epyT/pyT3SM 3SM3SM3SM 3SM6-8AC/SM 6-8AC/SM6-8AC/SM6-8AC/SM 6-8AC/SM21-4A
55MCP Process/ISMATEC SA/29.05.07/CB/GPepyT/pyTV2BSV3BS063083DA083082-orP182-orPxuanaC/slennahC2 3 1 1 1 1 1nellornepmuPstellaG/srelloR6 6 3 3 3 2 2py
56MCP Process/ISMATEC SA/29.05.07/CB/GP HinweisBei den Angaben in den nachfol-genden Tabellen handelt es sich nur um Richtwerte, die wie folgt
57MCP Process/ISMATEC SA/29.05.07/CB/GPOuverture du boîtier Afin de garantir l’étanchéité du boîtier, il est recommandé de n’ouvrir la pompe que pour
58MCP Process/ISMATEC SA/29.05.07/CB/GP Opening the casing hoodSealing the casing hoodWhen the casing hood fits tight, each single screw must be r
59MCP Process/ISMATEC SA/29.05.07/CB/GPRemplacement des fusiblesLes fusibles sont fixés sur le tab-leau de commande (dessus) et sur le tableau réseau (
6MCP Process/ISMATEC SA/29.05.07/CB/GP Warranty termsWe warrant the perfect functio-ning of our products, provided they have been installed and op
60MCP Process/ISMATEC SA/29.05.07/CB/GP MaintenanceProvided the MCP Process is ope-rated properly and in compliance with this manual, the tubing i
61MCP Process/ISMATEC SA/29.05.07/CB/GP HinweisBeachten Sie ebenfalls unsere Ga-rantie- und allgemeinen Verkaufs- und Lieferbedingungen.Bitte setzen
62MCP Process/ISMATEC SA/29.05.07/CB/GP
63MCP Process/ISMATEC SA/29.05.07/CB/GPRESETMINMINMAXRESETMINMINMAXMINMINMINCALMINCALMINCALMINCALMINCALMINCAL2x, or/oderOFFFlow rateVolumeFlow rateVol
64MCP Process/ISMATEC SA/29.05.07/CB/GP ISMATEC® Vertretung/Representative/ReprésentationISMATEC® - Ihr kompetenter Partner für anspruchsvolle Pu
7MCP Process/ISMATEC SA/29.05.07/CB/GPProduct Contents of the package MCP Process drive Magnet included Order No. ISM 915 incl. integrated power
8MCP Process/ISMATEC SA/29.05.07/CB/GP Panneau arrière1 Interrupteur principal2 Prise d’alimentation 3 RS232 IN (entrée femelle)4 RS232 OUT (s
9MCP Process/ISMATEC SA/29.05.07/CB/GPStarting the pump Mount the pump head accor-ding to the mounting instruc-tion manual supplied with the pump he